Agilent Technologies U1251A User Manual page 131

Handheld digital multimeter
Table of Contents

Advertisement

Table 5 Verification Test
Step
Test Function
1
Turn the rotary switch to the
V position
2
Press
frequency mode
3
Press
button to go to
Duty Cycle mode
Agilent U1251A/U1252A User's and Service Guide
Perform the verification test steps in the following Table 5:
Range
5 V
[1]
50 V
500 V
1000 V
button to go to
9.9999kHz
0.01% – 99.99%
Performance Tests and Calibration
5520A Output
Error from nominal 1 year
U1251A
5 V, 1 kHz
± 32.5 mV
5 V, 10 kHz
± 52.5 mV
5 V, 20 kHz
N/A
5 V, 30 kHz
± 84 mV
4.5 V, 100 kHz
N/A
50 V,1 kHz
± 325 mV
50 V,10 kHz
± 525 mV
50 V, 20 kHz
N/A
50 V, 30 kHz
± 840 mV
45 V, 100 kHz
N/A
500 V, 1 kHz
± 3.25 V
500 V, 10 kHz
± 5.25 V
1000 V, 1 kHz
± 10 V
0.48 V, 1 kHz
± 500 mHz
5.0 Vpp @
± 0.315%
50%, Square
Wave, 50 Hz
7
U1252A
± 22.5 mV
± 22.5 mV
± 41.5 mV
N/A
± 169.5 mV
± 225 mV
± 225 mV
± 415 mV
N/A
± 1.695 V
± 2.25 V
± 2.25V
± 8.0 V
± 500 mHz
± 0.315%