4.3.4 Electronics
L
WARNING:
Check the plug connectors for corrosion or damage before every ride,
since it could affect the efficiency of the electronics.
L
WARNING:
Remove the batteries before performing any maintenance, to avoid
involuntary conduction of currents.
Wipe the control unit with a cloth moistened by a few drops of a commercial domestic
cleaner. Do not use any abrasives or sharp-edged polishing equipment like a metal scrubber
or brush, since these can scratch the surface of the operator unit.
4.4

Inspection

In principle we recommend one inspection every year, and a minimum of one before usage is
resumed. All of the following checks must be performed and documented by authorized
persons:
Visually check the frame parts for plastic deformation, cracks and impaired
functioning (basic frame, seat frame, back frame, side parts, leg supports, motor
suspensions).
Visually check all plastic parts for cracks and brittle spots.
Check the solidity and seating of all screws.
Visually check all housings for damage, screws must be securely fixed, seals and
gaskets should not exhibit visible damage.
Visually check for damage to the paintwork (danger of corrosion).
Check the plug connectors for corrosion or damage, since it could affect the efficiency
of the electronics.
Check the operation of the tyres (free running, level rolling, axle play, tyres, profile,
condition of the rims, air pressure in the case of air tyres, floating axles, etc.).
Verify the amount of grease on the metal joints of movable parts.
Check the functioning of the arm supports and footrests (locking, load, deformation,
wear caused by loads).
Check the operation of other detachable parts (example: personal safety belt, battery
housing, etc.).
Check the cabling (especially for: crushing, abrasion, cuts, visible insulation of the
inner conductor, visible metallic veins, kinks, lumpiness, color changes of the outer
sleeve, brittle spots).
Electric leads to be securely placed to avoid shearing, crushing and other mechanical
stresses and strains.
For the charger of the electric wheelchair measure the resistance of the protective
braid (O) according to VDE 0702-1.
For the charger of the electric wheelchair measure the residual diversion current (A)
according to VDE 0702-1.
For the charger of the electric wheelchair measure insulation resistance (MO)
according to VDE 0702-1.
Check the functioning of the drives (during a test drive
etc.), if necessary: Measure the performance, first with no load and then with the
nominal load ("SWL"), to check for wear and tear of the motors by comparing the
values of the electric current with the values when the electric wheelchair was
delivered.
Check the condition of the batteries, covers and tubes.
Completeness of the delivery condition, instruction manual available?