Setting Inflation Pressure; Setting Measurement Intervals - Microlife WatchBP office Instruction Manual

Simultaneous dual-arm measurement and auscultatory function for accurate office blood pressure measurement
Hide thumbs Also See for WatchBP office:
Table of Contents

Advertisement

Special Function (cont.)
Setting inflation pressure in
«AUSCULTATION» Mode
Inflation pressure can be set manually from 140 to
280 mmHg at an intervals of 20 mmHg by pressing
and holding the Start/Stop Button for 7 seconds.
Press the M Button to adjust the pressure and press
Start/Stop again to save the setting.
AUSCULT
Press and hold
for 7 seconds...
Setting measuring intervals in
«ROUTINE» Mode
The default measuring intervals is 15 seconds. The
interval can be set as 15, 30, 45 or 60 seconds.
1) Pressing and holding the Start/Stop Button for 3
seconds.
2) Press the M Button to adjust the measurement
interval, then press the Start/Stop Button to
confirm, the meter will go back to stand by Mode.
15 30
60
45
Press and hold
for 3 seconds...
AUSCULT
26
EN