T
5.
Put now the hose of the sterile overflow Ø 8 mm into the overflow on the tank back and
connect these with the drain.
In order to ensure a perfect function of the sterile overflow, the gas cap
must be firmly locked.
09 Connect the tank supply cable to the 4-pin plug of the system.
09 Open the feedwater tap.
8. Check that all connections are secure and do not leak.
9. Check the feedwater pressure. It must be maintained within the specified pressure
range (see Technical specifications).
Before you switch the system on, read the procedure for rinsing reverse
osmosis membranes supplied in preservative solution in the „Rinsing and
disinfection" section!
10. Switch the system on at the on/off switch.
11. After a brief flush, your system produces purified water which flows into the tank.