Sessions > 4
7 Days:
5 Sessions
30 Days:
8 Sessions
System Leak
L/min
7 Day Avg:
0.0
30 Day Avg:
0.0
Apn-Hyp Index
7 Day Avg:
AHI = 0.0
30 Day Avg:
AHI = 0.0
4-6
3.
Sessions Greater than 4 Hours Screen
This screen displays the number of device therapy sessions that
exceeded 4 hours in 7-day and 30-day periods. The maximum value
for the 7-day period is 35 sessions, while the maximum value for the
30-day period is 150 sessions.
Note:
This screen is only for reference. Your home care provider
may periodically ask you for this information.
4.
System Leak Screen
This screen displays a 7-day and 30-day average of the leak history
for the device in liters per minute (LPM). System leak is a combina-
tion of intentional and unintentional air leak. Intentional leak is the
expected leak at the exhalation port. Some leak is required to mini-
mize CO
rebreathing. Unintentional leak occurs around the patient
2
interface. If you see a large increase in the amount of leak indicated
here, contact your home care provider for assistance.
Note:
This screen only displays if your home care provider has
enabled it. It is only for reference. Your home care provider
may periodically ask you for this information.
5.
Apnea/Hypopnea Index Screen
This screen displays your Apnea/Hypopnea index (AHI) for the last 7
and 30 days.
Note:
This screen only displays if your home care provider has
enabled it. It is only for reference. Your home care provider
may periodically ask you for this information.
6.
Compliance Check Value Screen
This screen displays a number used by your home care provider to
validate that the data provided by you is the data taken from the
Summary Data Screen. This is a "read-only" screen and you cannot
change the number on the screen.
Note: This screen is only for reference. Your home care provider
may periodically ask you for this information.
7. Summary Data Screen
This screen displays the total therapy hours for the device, the total
blower hours, and the total number of sessions greater than 4 hours
since the device was last reset by the home care provider. This is
a "read-only" screen and you cannot change the numbers on the
screen.
Note: This screen is only for reference. Your home care provider
may periodically ask you for this information.