IMPORTANT: Count forward 90 days from the date you
open a new bottle of AgaMatrix Control Solution. This is
your discard date. Write the discard date on the AgaMatrix
Control Solution bottle.
Important AgaMatrix Control Solution information:
[1] Use only AgaMatrix Control Solution with your AgaMatrix
WaveSense JAZZ System.
[2] Always replace the cap on the control solution bottle
immediately after use.
[3] Do not add water or any other liquid to control solution.
[4] Control solution tests should be performed within the system
operating temperature range of 10 °C to 40 °C (50 °F to 104 °F).
How to perform a control solution test:
[1] Inserting an AgaMatrix
WaveSense JAZZ Test Strip: Insert
a new AgaMatrix WaveSense JAZZ
Test Strip into the meter. The meter
will be activated when the test strip is
fully inserted.
[2] Meter Ready for Test: The meter now displays the
droplet symbol and the word "Fill". The AgaMatrix WaveSense
JAZZ System is now ready for you to apply control solution.
You do not need to set a calibration code for this meter.
18