[4] Viewing Control Solution Test Result:
The result will appear on the display and, if
the volume is enabled, your meter will beep
once. The date and the abbreviation "Ctrl" will
also appear on the display.
Compare the result of your control
solution test to the range printed on
the AgaMatrix WaveSense JAZZ Test
Strip vial label for the control solution
level (Normal or High) that you used.
Your control solution result should fall
within this range. The control solution
test result is stored in memory.
To deactivate the meter, remove the
used AgaMatrix WaveSense JAZZ
Test Strip. Dispose of used materials
properly.
If your control solution test results are
out of the range, repeat the test. If the
results from the control solution tests
continue to fall outside that range, do
not use the AgaMatrix WaveSense JAZZ System to test your
blood glucose. Call Customer Care.
IMPORTANT: Out-of-range results may be due to one or
more of the following factors:
[1] Expired or defective control solution.
[2] Expired or defective test strip.