Current-Input Fuses - Fluke 8845A Calibration Manual

Digital multimeter
Hide thumbs Also See for 8845A:
Table of Contents

Advertisement

3. Remove the fuse and replace with one rated appropriately for the selected line-power
voltage. See Table 2-1 for fuse ratings with specific line voltage.
4. Replace the selector block back into the fuse holder.
To avoid electric shock or fire, do not use makeshift fuses or
short-circuit the fuse holder. Use only Fluke fuses.
Line Voltage Selection
100
120
220
240

Current-Input Fuses

The 100 mA and 10 A inputs are protected by user-replaceable fuses.
The 100 mA input is protected by a fuse (A1F2) rated at 440 mA, 1000 V (fast
blow), 10,000 A minimum breaking capacity.
The 10 A input is protected by a fuse (A1F1) rated at 11 A, 1000 V (fast blow),
10,000 A minimum breaking capacity.
For protection against fire or arc flash, replace a blown fuse
only with one from Fluke.
XW Warning
Table 2-1. Line Voltage to Fuse Rating
Fuse Rating
0.25 A, 250 V (slow blow)
0.25 A, 250 V (slow blow)
0.125 A, 250 V (slow blow)
0.125 A, 250 V (slow blow)
Figure 2-1. Line Fuse Replacement
W Warning
General Maintenance
Fuse Replacement
Fluke Part No.
166306
166306
166488
166488
caw0201f.eps
2
2-5