Fluke 8845A Calibration Manual page 51

Digital multimeter
Hide thumbs Also See for 8845A:
Table of Contents

Advertisement

Table 3-6. 8845A/8846A AC Volts Frequency Verification Steps
Nominal Input
Ampl.
Freq.
1 V
10 Hz
1 V
40 Hz
100 mV
300 kHz
[1]
100 mV
1 MHz
[1] 8846A Only.
90-day Test Limits
High
Low
10.00300 Hz
9.99700 Hz
40.0040 Hz
39.9960 Hz
300.030 kHz
29.99700 kHz
1.000100 MHz
999.9000 kHz
Performance Test and Calibration
1-year Test Limits
High
10.00300 Hz
9.99700 Hz
40.01200 Hz
39.98800 Hz
300.0300 kHz
29.99700 kHz
1.0001000 MHz
999.9000 kHz
3
Performance Tests
Low
3-11

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

8846a

Table of Contents