Fluke 8845A Calibration Manual page 100

Digital multimeter
Hide thumbs Also See for 8845A:
Table of Contents

Advertisement

8845A/8846A
Calibration Manual
Nominal Output
Ampl.
0
10
-10
8845A Only DC Current
0
100.0E-6
-100.0E-6
0
1.0E-3
-1.0E-3
0
10.0E-3
-10.0E-3
0
100.0E-3
-100.0E-3
0
1
-1
0
1.9
-1.9
0
10
-10
8846A Only AC Current
100.0E-6
100.0E-6
100.0E-6
100.0E-6
1.0E-3
1.0E-3
A-14
Table A-2. Blank Verification Record for 1-Year Specifications (cont)
Range
Freq.
10
10
10
100.0E-6
100.0E-6
100.0E-6
1.0E-3
1.0E-3
1.0E-3
10.0E-3
10.0E-3
10.0E-3
100.0E-3
100.0E-3
100.0E-3
1
1
1
3
3
3
10
10
10
10
100.0E-6
1000
100.0E-6
5000
100.0E-6
10000
100.0E-6
10
1.0E-3
1000
1.0E-3
Test Limits
High
Low
800.0E-6
-800.0E-6
10.0158E+0
9.9842E+0
-9.9842E+0
-10.0158E+0
25.0E-9
-25.0E-9
100.075E-6
99.935E-6
-99.925E-6
-100.065E-6
50.0E-9
50.0E-9
1.00055E-3
999.45E-6
-999.45E-6
-1.00055E-3
2.0E-6
-2.0E-6
10.007E-3
9.993E-3
-9.993E-3
-10.007E-3
5.0E-6
-5.0E-6
100.055E-3
99.945E-3
-99.945E-3
-100.055E-3
200.0E-6
-200.0E-6
1.0012E+0
998.8E-3
-998.8E-3
-1.0012E+0
600.0E-6
-600.0E-6
1.90288E+0
1.89712E+0
-1.89712E+0
-1.90288E+0
800.0E-6
-800.0E-6
10.0158E+0
9.9842E+0
-9.9842E+0
-10.0158E+0
100.14E-6
99.86E-6
100.14E-6
99.86E-6
100.14E-6
99.86E-6
100.45E-6
99.55E-6
1.0014E-3
998.6E-6
1.0014E-3
998.6E-6
Results
Pass/Fail

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for: