8845A/8846A
Calibration Manual
Nominal Output
Amplitude
100.0 μA
[1]
100.0 μA
100.0 μA
[1]
100.0 μA
[1]
[1]
1.0 mA
1.0 mA
[1]
1.0 mA
[1]
1.0 mA
[1]
10.0 mA
10.0 mA
[1]
10.0 mA
[1]
10.0 mA
[1]
100.0 mA
100.0 mA
[1]
100.0 mA
[1]
100.0 mA
[1]
1 A
1 A
[1]
1 A
[1]
1 A
[1]
1.9 A
1.9 A
[1]
1.9 A
[1]
10 A
10 A
Notes:
[1] Optional test
3-24
Table 3-16. 8846A AC Current Verification Steps
Range
Freq.
100.0 μA
10 Hz
100.0 μA
1 kHz
100.0 μA
5 kHz
100.0 μA
10 kHz
10 Hz
1.0 mA
1 kHz
1.0 mA
5 kHz
1.0 mA
10 kHz
1.0 mA
10 Hz
10.0 mA
1 kHz
10.0 mA
5 kHz
10.0 mA
10 kHz
10.0 mA
10 Hz
100.0 mA
1 kHz
100.0 mA
5 kHz
100.0 mA
10 kHz
100.0 mA
45 Hz
1 A
1 kHz
1 A
5 kHz
1 A
10 kHz
1 A
45 Hz
3 A
1 kHz
3 A
10 kHz
3 A
45 Hz
10 A
1 kHz
10 A
90-day Test Limits
High
Low
100.14 μA
99.86 μA
100.14 μA
99.86 μA
100.14 μA
99.86 μA
100.45 μA
99.55 μA
998.6 μA
1.0014 mA
998.6 μA
1.0014 mA
998.6 μA
1.0014 mA
995.5 μA
1.0045 mA
10.014 mA
9.986 mA
10.014 mA
9.986 mA
10.014 mA
9.986 mA
10.045 mA
9.955 mA
100.14 mA
99.86 mA
100.14 mA
99.86 mA
100.14 mA
99.86 mA
100.45 mA
99.55 mA
1.0014 A
998.6 mA
1.0014 A
998.6 mA
1.0014 A
998.6 mA
1.0105 A
989.5 mA
1.90465 A
1.89535 A
1.90465 A
1.89630 A
1.92765 A
1.87235 A
10.02100 A
9.97900 A
10.02100 A
9.97900 A
1-year Test Limits
High
Low
100.14 μA
99.86 μA
100.14 μA
99.86 μA
100.14 μA
99.86 μA
100.45 μA
99.55 μA
998.6 μA
1.0014 mA
998.6 μA
1.0014 mA
998.6 μA
1.0014 mA
995.5 μA
1.0045 mA
10.014 mA
9.986 mA
10.014 mA
9.986 mA
10.014 mA
9.986 mA
10.045 mA
9.955 mA
100.14 mA
99.86 mA
100.14 mA
99.86 mA
100.14 mA
99.86 mA
100.45 mA
99.55 mA
1.00140 A
998.6 mA
1.00140 A
998.6 mA
1.00140 A
998.6 mA
1.01050 A
989.5 mA
1.90465 A
1.89535 A
1.90465 A
1.89535 A
1.92765 A
1.87235 A
10.02100 A
9.97900 A
10.02100 A
9.97900 A