Table 3-18. 8845A/8846A Adjustment Steps (cont)
Step
Modes
Value Range
75
GVDC
10
76
GVDC
-10
77
GVDC
1
78
GVDC
-1
79
GVDC
0.1
80
GVDC
-0.1
Hi IDC
Gain
81
GIDC
1
82
GIDC
-1
83
GIDC
10
84
GIDC
-10
Hi IAC
Gain
85
GIAC
10
86
GIACS
10
87
GIAC
1
88
GIACS
1
Low IAC
Gain
89
GIAC
100.0E-3
90
GIACS
100.0E-3
91
GIAC
10.0E-3
92
GIACS
10.0E-3
93
GIAC
1.0E-3
94
GIACS
1.0E-3
95
GIAC
100.0E-6
96
GIACS
100.0E-6
Lo IDC
Gain
97
GIDC
100.0E-6
98
GIDC
-100.0E-6
99
GIDC
1.0E-3
100
GIDC
-1.0E-3
Input Signal
10
-10
1
-1
0.1
-0.1
1
-1
10
-10
10
10
1
1
100.0E-3
100.0E-3
10.0E-3
10.0E-3
1.0E-3
1.0E-3
100.0E-6
100.0E-6
100.0E-6
-100.0E-6
1.0E-3
-1.0E-3
Performance Test and Calibration
Adjustment (Calibration)
Description
DC 10 V
DC 10 V
DC 1 V
DC 1 V
DC 100 mV
DC 100 mV
DC 1 A
DC 1 A
DC 10 A
DC 10 A
AC 10 A
AC 10 A
AC 1 A
AC 1 A
AC 100 mA
AC 100 mA
AC 10 mA
AC 10 mA
AC 1 mA
AC 1 mA
AC 100 uA
AC 100 uA
DC 100 uA
DC 100 uA
DC 1 mA
DC 1 mA
3
Series
N
N
N
N
N
N
N
N
N
N
Y
N
Y
N
Y
N
Y
N
Y
N
N
N
N
N
N