Nominal
Input (V)
0
[1]
100.0 mV
[1]
-100.0 mV
0 V
[1]
1 V
[1]
-1 V
0 V
3-6
INPUT
SENSE
V
4 W
2W/4W
HI
HI
1000 V CAT I
FUSED
300V
600V CAT II
LO
LO
1V
FUSED
100
10A
mA
REAR
FRONT
5520A
CALIBRATOR
NORMAL
AUX
V, , ,RTD
A, -SENSE, AUX V
HI
FUSED
LO
GUARD
20A
TC
20V PK MAX
20V PK MAX
UUT
A.
EX SNS
INPUT
SENSE
V
4 W
2W/4W
HI
HI
1000 V CAT I
300V
600V CAT II
LO
LO
1V
FUSED
100
10A
mA
REAR
FRONT
NC
Figure 3-1. DC Volts Test Equipment Setup with 5520A
Table 3-2. 8846A DC Volts Verification Steps
90-day Test Limits
Range
High
3.5 μV
0.100
0.100
100.006 mV
0.100
-99.994 Mv
7.0 μV
1
1
1.000025 V
1
-0.999975 V
50.0 μV
10
8846A
6-1/2 DIGIT PRECISION MULTIMETER
+1.000002
V
BACK
F1
F2
F3
F4
DC V
ACV
DC I
AC I
TRIG
RANGE
FREQ
TEMP
ZERO
ANALYZE
PERIOD
SCOPE
OUT
STBY
EARTH
SCOPE
OPR
μ
7
8
9
TRIG
n
4
5
6
p
1
2
3
/ +
0
SHIFT
: OFF
: OFF
EX GRD
Low
-3.5 μV
3.5 μV
99.994 mV
100.0072 mV
-100.006 mV
-99.9928 mV
-7.0 μV
7.0 μV
0.999975 V
1.000032 V
-1.000025 V
-0.999968 V
-50.0 μV
50.0 μV
F5
MEMORY
INSTR
SETUP
MEAS
SETUP
PREV
EXGRD
MENU
dBm
sec
m
V
Hz
W
F
k
A
C
M
F
ENTER
5720A Calibrator
OUTPUT
SENSE
V A
V
WIDEBAND
HI
HI
LO
LO
HI
AUX
GUARD
GROUND
CURRENT
1-year Test Limits
High
Low
-3.5 μV
99.9928 mV
-100.0072 mV
-7.0 μV
0.999968 V
-1.000032 V
-50.0 μV