8808A
Getting Started Manual
Current-Input Fuses
The 200 mA and 10 A inputs are protected by user-replaceable fuses.
For protection against fire or arc flash, replace a blown fuse
with a fuse of an identical rating.
To test the current-input fuses:
6
Table 3. Line Voltage to Fuse Rating
Line Voltage Selection
100 / 120
220 / 240
Figure 1. Replacing the Line-Power Fuse
•
The 200 mA input is protected by a fuse rated at 440 mA, 1000 V (fast blow),
10,000 A minimum breaking capacity.
•
The 10 A input is protected by a fuse rated at 11 A, 1000 V (fast blow), 10,000 A
minimum breaking capacity.
1. Turn on the Meter and plug a test lead into the INPUT VZYR HI terminal.
2. Press O.
3. Press Vto set the range to 200 Ω. Only the 200 Ω, 2 kΩ, and 20 kΩ ranges can
be used to test the mA input fuse.
4. Insert the other end of the test lead into the mA terminal. If the fuse is good, the
Meter displays a reading of 0.000 MΩ. If the fuse is blown, the Meter displays 0L
to indicate an overload.
5. Remove the test lead from the mA terminal and insert it into the 10 A terminal.
If the fuse is good, the Meter displays a reading of <1.000 Ω. If the fuse is blown,
the Meter displays 0L to indicate an overload.
0.125 A, 250 V (slow blow)
0.063 A, 250 V (slow blow)
XW Warning
Fuse Rating
eue20.eps