8845A/8846A
Calibration Manual
B-8
PRINT @<address of meter>, "CAL? ON"
INPUT LINE @<address of meter>, A$
### Calibrate 0.1 A AC @1200 Hz
# 0.1 A AC
PRINT @<address of meter>, "CAL:VAL GIAC,100.0E-3"
PRINT @<address of meter>, "CAL? ON"
INPUT LINE @<address of meter>, A$
### Calibrate 0.1 A AC @1200 Hz
# 0.1 A AC
PRINT @<address of meter>, "CAL:VAL GIACS,100.0E-3"
PRINT @<address of meter>, "CAL? ON"
INPUT LINE @<address of meter>, A$
### Calibrate 0.01 A AC @1200 Hz
# 0.01 A AC
PRINT @<address of meter>, "CAL:VAL GIAC,10.0E-3"
PRINT @<address of meter>, "CAL? ON"
INPUT LINE @<address of meter>, A$
### Calibrate 0.01 A AC @1200 Hz
# 0.01 A AC
PRINT @<address of meter>, "CAL:VAL GIACS,10.0E-3"
PRINT @<address of meter>, "CAL? ON"
INPUT LINE @<address of meter>, A$
### Calibrate 0.001 A AC @1200 Hz 8846 only
# 0.001 A AC
PRINT @<address of meter>, "CAL:VAL GIAC,1.0E-3"
PRINT @<address of meter>, "CAL? ON"
INPUT LINE @<address of meter>, A$
### Calibrate 0.001 A AC @1200 Hz 8846 only
# 0.001 A AC
PRINT @<address of meter>, "CAL:VAL GIACS,1.0E-3"
PRINT @<address of meter>, "CAL? ON"
INPUT LINE @<address of meter>, A$
### Calibrate -0.0001 A AC @1200 Hz 8846 only
# 0.0001 A AC
PRINT @<address of meter>, "CAL:VAL GIAC,100.0E-6"
PRINT @<address of meter>, "CAL? ON"
INPUT LINE @<address of meter>, A$
### Calibrate 0.0001 A AC @1200 Hz 8846 only
# 0.0001 A AC
PRINT @<address of meter>, "CAL:VAL GIACS,100.0E-6"
PRINT @<address of meter>, "CAL? ON"
INPUT LINE @<address of meter>, A$
PRINT @<address of meter>, "CAL:REC"
### Calibrate 0.0001 ADC
# 0.0001 ADC
PRINT @<address of meter>, "CAL:VAL GIDC,100.0E-6"
PRINT @<address of meter>, "CAL? ON"
INPUT LINE @<address of meter>, A$
### Calibrate -0.0001 ADC
# -0.0001 ADC
PRINT @<address of meter>, "CAL:VAL GIDC,-100.0E-6"
PRINT @<address of meter>, "CAL? ON"
INPUT LINE @<address of meter>, A$
### Calibrate 0.001 ADC
# 0.001 ADC