Fluke 8845A Calibration Manual page 70

Digital multimeter
Hide thumbs Also See for 8845A:
Table of Contents

Advertisement

8845A/8846A
Calibration Manual
Step
48
49
AC
50
51
52
53
54
ACV
Gain
55
56
57
58
59
60
61
62
63
64
65
66
67
68
69
70
VDC
Gain
71
72
73
74
3-30
Table 3-18. 8845A/8846A Adjustment Steps (cont)
Modes
Value Range
ZIACS
1
ZIAC
10
ZIACS
10
Linearity
ACLIN
1.19
ACLIN
0.8
ACLIN
0.4
ACLIN
0.005
GVAC
0.1
GVACS
0.1
ACPOLE
0.1
GVAC
1
GVACS
1
FVAC
1
ACPOLE
1
GVAC
10
GVACS
10
ACPOLE
10
GVAC
100
GVACS
100
ACPOLE
100
GVAC
1000
GVACS
1000
ACPOLE
1000
GVDC
1000
GVDC
-1000
GVDC
100
GVDC
-100
Input Signal
10 A to Lo short
10 A to Lo short
10 A to Lo short
1.19 @1200 Hz
0.8 @1200 Hz
0.4 @1200 Hz
0.005 @1200 Hz
0.1 @1200 Hz
0.1 @1200 Hz
0.1 @50000 Hz
1 @1200 Hz
1 @1000 Hz
1 @10 Hz
1 @50000 Hz
10 @1200 Hz
10 @1200 Hz
10 @50000 Hz
100 @1200 Hz
100 @1200 Hz
100 @50000 Hz
1000 @1200 Hz
1000 @1200 Hz
329 @50000 Hz
1000
-1000
100
-100
Description
AC 1 A
AC 10 A
AC 10 A
AC 1 V
AC 1 V
AC 1 V
AC 1 V
AC 100 mV
AC 100 mV
A 100 mV
AC 1 V
AC 1 V
AC 1 V
AC 1 V
AC 10 V
AC 10 V
AC 10 V
AC 100 V
AC 100 V
AC 100 V
AC 1000 V (8845A use
750V on 750V range)
AC 1000 V (8845A use
750V on 750V range)
AC 1000 V (use 8845A
750V range)
DC 1000 V
DC 1000 V
DC 100 V
DC 100 V
Series
Y
Y
N
N
N
N
N
Y
N
N
Y
N
N
N
Y
N
N
Y
N
N
Y
N
N
N
N
N
N

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

8846a

Table of Contents