8845A/8846A
Calibration Manual
Step
101
102
103
104
Ω Gain
105
106
107
108
109
110
111
112
Misc
Gain
113
114
115
Once familiar with the calibration series of setups, the calibration time may be sped up by
using the command "CAL OFF". This command allows the instrument to automatically
go to the next logical step in the series. The last column in Table 3-18 identifies which
steps in the series can be run automatically. For example, all of the Open steps can be run
by entering:
Another example is automatically running the ACV zeros from step 3 through step 12 in
Table 3-18 with:
Using the CAL? command without an argument turns off the single step
feature.
AC linearity steps must be run with AC Gain steps to complete the AC
calibration.
3-32
Table 3-18. 8845A/8846A Adjustment Steps (cont)
Modes
Value Range
GIDC
10.0E-3
GIDC
-10.0E-3
GIDC
100.0E-3
GIDC
-100.0E-3
GRES
100000000
GRES
10000000
GRES
1000000
GRES
100000
GRES
10000
GRES
1000
GRES
100
GRES
10
GRES
1000000000
GCAP1
10.0E-9
GCAP2
10.0E-9
CAL:CAL ORES, 100000000
CAL? OFF
CAL:CAL ZVAC, 100.0E-3
CAL? OFF
Input Signal
10.0E-3
-10.0E-3
100.0E-3
-100.0E-3
100000000
10000000
1000000
100000
10000
1000
100
10
1000000000
10.0E-9
10.0E-9
Notes
Description
DC 10 mA
DC 10 mA
DC 100 mA
DC 100 mA
R 100M Ω
4W 10M Ω
4W 1M Ω
4W 100 kΩ
4W 10 kΩ
4W 1 kΩ
4W 100 Ω
4W 10 Ω (8846A only)
R1G Ω (8846A only)
C10NF (8846A only)
C10NF (8846A only)
Series
N
N
N
N
N
N
N
N
N
N
N
N
N
Y
N