Fluke 8845A Calibration Manual page 50

Digital multimeter
Hide thumbs Also See for 8845A:
Table of Contents

Advertisement

8845A/8846A
Calibration Manual
Nominal Input
Ampl.
[1]
100.0 mV
100.0 mV
100.0 mV
100.0 mV
100.0 mV
[1]
1 V
1 V
1 V
1 V
1 V
[1]
10 V
[1]
10 V
10 V
10 V
3 V
100 V
[1]
100 V
100 V
[1]
100 V
750 V
750 V
750 V
320 V
320 V
320 V
[1] 5520A must be used with 8508A to obtain suitable test uncertainty ratio.
3-10
Table 3-5. 8845A AC Volts Verification Steps
Range
Freq.
10 Hz
0.100
20 kHz
0.100
50 kHz
0.100
100 kHz
0.100
300 kHz
0.100
10 Hz
1
20 kHz
1
50 kHz
1
100 kHz
1
300 kHz
1
10 Hz
10
20 kHz
10
50 kHz
10
100 kHz
10
300 kHz
10
45 Hz
100
20 kHz
100
50 kHz
100
100 kHz
100
45 Hz
750
1 kHz
750
10 kHz
750
20 kHz
750
50 kHz
750
100 kHz
750
90-day Test Limits
High
Low
100.09 mV
99.91 mV
100.09 mV
99.91 mV
100.16 mV
99.84 mV
100.68 mV
99.32 mV
104.5 mV
95.5 mV
1.0008 V
999.2 mV
1.0008 V
999.2 mV
1.0016 V
998.4 mV
1.0068 V
993.2 mV
1.045 V
955.0 mV
10.008 V
9.992 V
10.008 V
9.992 V
10.016 V
9.984 V
10.068 V
9.932 V
3.17 V
2.83 V
100.08 V
99.92 V
100.08 V
99.92 V
100.16 V
99.84 V
100.68 V
99.32 V
750.600
749.400
750.600
749.400
750.600
749.400
320.385
319.615
320.727 V
319.273 V
322.52 V
317.48 V
1-year Test Limits
High
Low
100.1 mV
99.9 mV
100.1 mV
99.9 mV
100.17 mV
99.83 mV
100.68 mV
99.32 mV
104.5 mV
95.5 mV
1.0009 V
999.1 mV
1.0009 V
999.1 mV
1.0017 V
998.3 mV
1.0068 V
993.2 mV
1.045 V
955.0 mV
10.009 V
9.991 V
10.009 V
9.991 V
10.017 V
9.983 V
10.068 V
9.932 V
3.17 V
2.83 V
100.09 V
99.91 V
100.09 V
99.91 V
100.17 V
99.83 V
100.68 V
99.32 V
750.675 V
749.325 V
750.675 V
749.325 V
750.675 V
749.325 V
320.417 V
319.583 V
320.759 V
319.241 V
322.52 V
317.48 V