Adjustment (Calibration) - Fluke 8845A Calibration Manual

Digital multimeter
Hide thumbs Also See for 8845A:
Table of Contents

Advertisement

Nominal Output
Ampl
Freq
[1]
10.0 mA
10 Hz
10.0 mA
1 kHz
[1]
10.0 mA
5 kHz
[1]
10.0 mA
10 kHz
[1]
100.0 mA
10 Hz
100.0 mA
1 kHz
[1]
100.0 mA
5 kHz
[1]
100.0 mA
10 kHz
[1]
1 A
45 Hz
1 A
1 kHz
[1]
1 A
5 kHz
[1]
1 A
10 kHz
[1]
1.9 A
45 Hz
1.9 A
1 kHz
[1]
1.9 A
10 kHz
[1]
10 A
45 Hz
10 A
1 kHz
Notes:
[1] Optional test

Adjustment (Calibration)

Meter adjustments, or calibration, should be performed at the desired interval, or
whenever a verification test indicates a Meter function is out of tolerance. The Meter
accuracy will stay within specifications only if the adjustment procedure is performed at
regular intervals. A one-year interval is adequate for most applications. The Meter's
accuracy specifications are not valid beyond the one-year interval.
Meter adjustments are accessed only through the remote interface with a series of
adjustment steps. The remote program directs the test equipment to apply a series of
shorts, opens, voltages, currents, and capacitance (8846A only) to the Meter. At each
step, the Meter internally makes the necessary adjustment to bring the Meter into
specification. No internal mechanical adjustments are necessary.
Using an automated, computer-controlled procedure, the calibration and verification
procedures can be performed on the Meter in under 60 minutes. A sample adjustment
program is listed on the "Sample Adjustment Program" section later in this manual. A
MetCal program is available at
The Meter's adjustments are password protected to prevent accidental or unauthorized
adjustments. The security password must be entered through the front panel or remote
interface before adjustments can be made to the Meter.
Table 3-17. 8845A AC Current Verification Steps
90-day Test Limits
Range
High
10.0 mA
10.014 mA
10.0 mA
10.014 mA
10.0 mA
10.014 mA
10.0 mA
10.045 mA
100.0 mA
100.14 mA
100.0 mA
100.14 mA
100.0 mA
100.14 mA
100.0 mA
100.45 mA
1 A
1.0014 A
1 A
1.0014 A
1 A
1.0014 A
1 A
1.0105 A
3 A
1.90465 A
3 A
1.90465 A
3 A
1.92765 A
10 A
10.01900 A
10 A
10.01900 A
www.fluke.com
Performance Test and Calibration
1-year Test Limits
Low
High
9.986 mA
10.014 mA
9.986 mA
10.014 mA
9.986 mA
10.014 mA
9.955 mA
10.045 mA
99.86 mA
100.14 mA
99.86 mA
100.14 mA
99.86 mA
100.14 mA
99.55 mA
100.45 mA
998.6 mA
1.00140 A
998.6 mA
1.00140 A
998.6 mA
1.00140 A
989.5 mA
1.01050 A
1.89535
1.90465 A
1.89630
1.90465 A
1.87235
1.92765 A
9.98100 A
10.02100 A
9.98100 A
10.02100 A
to adjust the Meter.

Adjustment (Calibration)

Low
9.986 mA
9.986 mA
9.986 mA
9.955 mA
99.86 mA
99.86 mA
99.86 mA
99.55 mA
998.6 mA
998.6 mA
998.6 mA
989.5 mA
1.89535 A
1.89535 A
1.87235 A
9.97900 A
9.97900 A
3