Fluke 8845A Calibration Manual page 68

Digital multimeter
Hide thumbs Also See for 8845A:
Table of Contents

Advertisement

8845A/8846A
Calibration Manual
Step
Open
0
1
2
ACV
Zero
3
4
5
6
7
8
9
10
11
12
DCV
Zero
13
14
15
16
17
18
Ohm
Zero
19
20
21
22
23
24
3-28
Table 3-18. 8845A/8846A Adjustment Steps
Modes
Value Range
ORES
100000000
ORES
1000000000
ZCAP
1.00E-09
ZVAC
100.0E-3
ZVACS
100.0E-3
ZVAC
1
ZVACS
1
ZVAC
10
ZVACS
10
ZVAC
100
ZVACS
100
ZVAC
1000
ZVACS
1000
ZVDC
1000
ZVDC
100
ZVDC
10
ZVDC
1
ZVDC
0.1
DFVDC
0.1
ZRES
10000000
ZRES
1000000
ZRES
100000
ZRES
10000
ZRES
1000
ZRES
100
Input Signal
open
open
open
4-wire low-thermal short
4-wire low-thermal short
4-wire low-thermal short
4-wire low-thermal short
4-wire low-thermal short
4-wire low-thermal short
4-wire low-thermal short
4-wire low-thermal short
4-wire low-thermal short
4-wire low-thermal short
4-wire low-thermal short
4-wire low-thermal short
4-wire low-thermal short
4-wire low-thermal short
4-wire low-thermal short
4-wire low-thermal short
4-wire low-thermal short
4-wire low-thermal short
4-wire low-thermal short
4-wire low-thermal short
4-wire low-thermal short
4-wire low-thermal short
Description
OHM 100M open
terminals
OHM 1G open terminals
(8846A only)
CAP 1 nF open
terminals (8846A only)
AC 100 mV
AC 100 mV
AC 1V
AC 1V
AC 10V
AC 10V
AC 100V
AC 100V
AC 1000V
AC 1000V
DC 1000V
DC 100V
DC 10V
DC 1V
DC 100mV
DC 100mV
4W Ohm 10 MOHM
4W 1 MOHM
4W 100 kOHM
4W 10 kOHM
4W 1 kOHM
4W 100 OHM
Series
Y
Y
N
Y
Y
Y
Y
Y
Y
Y
Y
Y
N
Y
Y
Y
Y
N
N
Y
Y
Y
Y
Y
Y

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

8846a

Table of Contents