Fluke 8845A Calibration Manual page 92

Digital multimeter
Hide thumbs Also See for 8845A:
Table of Contents

Advertisement

8845A/8846A
Calibration Manual
Nominal Output
Ampl.
8846A Capacitance
0
1.0E-9
10.0E-9
100.0E-9
1.0E-6
10.0E-6
100.0E-6
1.0E-3
10.0E-3
100.0E-3
8846A DC Current
0
100.0E-6
-100.0E-6
0
1.0E-3
-1.0E-3
0
10.0E-3
-10.0E-3
0
100.0E-3
-100.0E-3
0
1
-1
0
1.9
-1.9
A-6
Table A-1. Blank Verification Record for 90-Day Specifications (cont)
Range
Freq.
1.0E-9
25.0E-12
1.0E-9
1.045E-9
10.0E-9
10.15E-9
100.0E-9
101.5E-9
1.0E-6
1.015E-6
10.0E-6
10.15E-6
100.0E-6
101.5E-6
1.0E-3
1.015E-3
10.0E-3
10.15E-3
100.0E-3
104.2E-3
100.0E-6
25.0E-9
100.0E-6
100.065E-6
100.0E-6
-99.935E-6
1.0E-3
50.0E-9
1.0E-3
1.00035E-3
1.0E-3
-999.65E-6
10.0E-3
2.0E-6
10.0E-3
10.005E-3
10.0E-3
-9.995E-3
100.0E-3
5.0E-6
100.0E-3
100.035E-3
100.0E-3
-99.965E-3
1
200.0E-6
1
1.0006E+0
1
-999.4E-3
3
600.0E-6
3
1.90212E+0
3
-1.89788E+0
Test Limits
High
Low
0
955.0E-12
9.85E-9
98.5E-9
985.0E-9
9.85E-6
98.5E-6
985.0E-6
9.85E-3
95.8E-3
-25.0E-9
99.935E-6
-100.065E-6
-50.0E-9
999.65E-6
-1.00035E-3
-2.0E-6
9.995E-3
-10.005E-3
-5.0E-6
99.965E-3
-100.035E-3
-200.0E-6
999.4E-3
-1.0006E+0
-600.0E-6
1.89788E+0
-1.90212E+0
Results
Pass/Fail