GENERAL SPECIFICATIONS AND STANDARDS COMPLIANCE
POWER
Voltage: 90–134V and 180–250V, universal self-selecting.
Frequency: 50Hz, 60Hz, or 400Hz, self-identifying at power-up.
Consumption: <55VA.
ENVIRONMENTAL
Operating Temperature: 0°C to 50°C.
Storage Temperature: –40°C to 70°C.
Humidity: 80% R.H., 0°C to 35°C, per MIL-T-28800E
CALIBRATION
Type: Software. No manual adjustments required.
Sources: 2 DC voltages, 6 resistances, and 5 DC currents. All other functions calibrated (adjusted) from these sources and a short circuit. No
AC calibrator required for adjustment.
Average Time to Perform: 40 minutes for comprehensive calibration, 6 minutes for AC-only calibration.
PROCESS: MIL-STD 45662A.
PHYSICAL
Case Dimensions: 90mm high × 214mm wide × 369mm deep (3
Working Dimensions: From front of case to rear including power cord and IEEE-488 connector: 15.0 inches.
Net Weight: <4.2kg (<9.2 lbs.).
Shipping Weight: <9.1kg (<20lbs.).
STANDARDS
EMI/RFI: Conforms to VDE 0871B (per Vfg 1046/1984), IEC 801-2. Meets FCC part 15 Class B, CISPR-22 (EN55022).
Safety: Conforms to IEC348, CAN/CSA-C22.2. No. 231, MIL-T-28800E
ACCESSORIES SUPPLIED
The unit is shipped with line cord, high performance modular test leads, manuals, option slot cover, and full calibration data.
Note 1: For MIL-T-28800E, applies to Type III, Class 5, Style E.
1
Para 4.5.5.1.2.
in. × 8
in. × 14
1
. Designed to UL1244.