5 Monitoring Contractions (External)
Symptom
Worse
trace
quality
fluctuant TOCO baseline
Too high TOCO sensitivity
(higher than 100 unit)

Attachment 3 Monitoring Figure

Possible Cause
or
Too tight or too loose
abdomen
belt or no
elasticity
Maternal Movement
Fetal Movement
The body pressure from
uterus
to
transducer is far higher
than the average value.
Ensure the abdomen belt has been
used accurately and neither too
tight, nor too loose
Relax patient's spirit
Wait for a moment then monitor
Insure favorable contact for patient
TOCO
skin with TOCO transducer. Change
the position of TOCO transducer, if
necessary.
Solution