Contec CMS-60C Instructions Manual
Hide thumbs Also See for CMS-60C:

Advertisement

Quick Links

Instructions for CMS-60C Pulse Oximeter
Dear users, thank you very much for purchasing the Pulse Oximeter.
This Manual is written and compiled in accordance with the council directive MDD93/42/EEC for
medical devices and harmonized standards. In case of modifications and software upgrades, the
information contained in this document is subject to change without notice.
The Manual describes, in accordance with the Pulse Oximeter's features and requirements, main
structure, functions, specifications, correct methods for transportation, installation, usage, operation,
repair, maintenance and storage, etc. as well as the safety procedures to protect both the user and
equipment. Refer to the respective chapters for details.
Please read the User Manual carefully before using this product. The User Manual which describes
the operating procedures should be followed strictly. Failure to follow the User Manual may cause
measuring abnormality, equipment damage and human injury. The manufacturer is NOT responsible
for the safety, reliability and performance issues and any monitoring abnormality, human injury and
equipment damage due to users' negligence of the operation instructions. The manufacturer's
warranty service does not cover such faults.
Owing to the forthcoming renovation, the specific products you received may not be totally in accor
dance with the description of this User Manual. We would sincerely regret for that.
This product is medical device, which can be used repeatedly.
WARNING:
 Uncomfortable or painful feeling may appear if using the device ceaselessly, especially for
the microcirculation barrier patients. It is recommended that the sensor should not be
applied to the same finger for over 2 hours.
 For the special patients, there should be a more prudent inspecting in the placing process.
The device cannot be clipped on the edema and tender tissue.
 The light (the infrared is invisible) emitted from the device is harmful to the eyes, so the
user and the maintenance man should not stare at the light.
 Testee cannot use enamel or other makeup.
 Testee's fingernail cannot be too long.
 Please refer to the correlative literature about the clinical restrictions and caution.
 This device is not intended for treatment.
The User Manual is published by our company. All rights reserved.
CMS-60C
Medical Test Supply

Advertisement

Table of Contents
loading

Summary of Contents for Contec CMS-60C

  • Page 1 Medical Test Supply Instructions for CMS-60C Pulse Oximeter Dear users, thank you very much for purchasing the Pulse Oximeter. This Manual is written and compiled in accordance with the council directive MDD93/42/EEC for medical devices and harmonized standards. In case of modifications and software upgrades, the information contained in this document is subject to change without notice.
  • Page 2: Table Of Contents

    7 MAINTAIN, TRANSPORTATION AND STORAGE ..............11 7.1 C ......................11 LEANING AND ISINFECTING 7.2 M ............................11 AINTAIN 7.3 T ......................11 RANSPORTATION AND STORAGE 8 TROUBLESHOOTING ........................11 9 KEY OF SYMBOLS .......................... 12 10 FUNCTION SPECIFICATION ..................... 13 CMS-60C...
  • Page 3 Medical Test Supply APPENDIX ............................14 CMS-60C...
  • Page 4: Safety

    Refer to User Manual in the relative chapter (7.1)for instructions of cleaning and disinfection.  Do not have the oximeter immerged in liquid. When it needs cleaning, please wipe its surface with medical alcohol by soft material. Do not spray any liquid on the device CMS-60C...
  • Page 5: Overview

    2.1 Features A. Operation of the product is simple and convenient. B. The product is small in volume, light in weight and convenient in carrying. CMS-60C...
  • Page 6: Major Applications And Scope Of Application

    Then measured signal can be obtained by a photosensitive element, information acquired through which will be shown on screen through treatment in electronic circuits and microprocessor. Figure 1. CMS-60C...
  • Page 7: Technical Specifications

    ±1%. F. Power supply requirement : 3.6 V DC ~ 4.2V DC. G. Optical Sensor Red light (wavelength is 660nm , 6.65mW) Infrared (wavelength is 880nm, 6.75mW) H. Adjustable alarm range : 0%~100% Pulse Rate : 0bpm~254bpm CMS-60C...
  • Page 8: Installation

    2. USB port :It is used to connect a personal computer to export the trend data(or real-time data) or charge the lithium battery via a data line. 5.3 RearView Finger 4. RearView (Refer to Figure 4 and insert the lithium battery properly in the right direction.) CMS-60C...
  • Page 9: Accessories

    。 Intermittent alarm will occur and the user interface presents "FINGER OUT". Medium priority indicating that prompt operator response is required. Figure 5 ( Actual probe may be different with the probe as figure 5,please accept the actual probe with the device ) B. Change display direction: CMS-60C...
  • Page 10 If the alarm function is on, the device will provide medium-priority alarm signal when the data of SpO or pulse rate is beyond the limit. Intermittent alarm will occur and the measurement shows in yellow. CMS-60C...
  • Page 11 9:if it is in recording s tate,the"Stop"is displayed on menu;if it is not in recording state,the"Start"is displayed on menu. Figure 9. b. When the"Start"is displayed on the menu,if the memory still have the former data in storage ,the CMS-60C...
  • Page 12 Please connect the device with computer by the USB data line , then double click "SpO2 Assistant "i con to run the PC software.The functions such as uploading data and change device ID could be carried out by the software. Please refer to <SpO Assistant user manual> for detail. CMS-60C...
  • Page 13: Attention For Operation

    SpO and pulse rate. F. Excessive ambient light may affect the measuring result. It includes fluorescent lamp, dual ruby light, infrared heater, direct sunlight and etc. CMS-60C...
  • Page 14: Clinical Restrictions

    A. The packed device can be transported by ordinary conveyance or according to transport contract. The device cannot be transported mixed with toxic, harmful, corrosive material. B. The packed device should be stored in room with no corrosive gases and good ventilation. Temperature: -40°C~60°C; Humidity: ≤95% 8 Troubleshooting Trouble Possible Reason Solution CMS-60C...
  • Page 15: Key Of Symbols

    9 Key of Symbols Signal Description Warning – See User Manual The pulse oxygen saturation(%) PRbpm Pulse rate (bpm) Close the alarm sound indication forever Pause the alarm sound indication Open the alarm sound indication Close the pulse sound indication CMS-60C...
  • Page 16: Function Specification

    TFT display Parameter Specification Measuring range 0%~100%, (the resolution is 1%). Accuracy 70%~100%:±2% ,Below 70% unspecified. Average value Calculate the Average value in every 4 measure value. The deviation between average value and true value does not exceed 1%. CMS-60C...
  • Page 17 Output power 1.25 W Oximeter Probe Wavelength:660nm 880nm Dimensions and Weight Dimensions 87(L) × 45(W) × 22(H) mm Weight About 175 g (with the lithium battery*1) Appendix State Alarm condition delay Alarm signal generation delay Low voltage alarm 20ms CMS-60C...
  • Page 18 Medical Test Supply alarm 330ms 20ms Pulse rate alarm 330ms 20ms Probe error alarm 16ms 20ms CMS-60C...