Page 1
CMS-50E Pulse Oximeter CONTEC MEDICAL SYSTEMS CO., LTD...
Page 2
This Manual is written and complied in accordance with the council directive MDD 93/42/EEC for medical devices and corresponding standards. The Manual written is in accordance the Fingertip Pulse Oximeter. In case of modifications and software upgrades, you will be advised in due time with a Modification Notice.
This device is not intended for treatment. Table of Contents 1. SAFETY .............................1 1.1. I ................... 1 NSTRUCTIONS FOR SAFE OPERATIONS 1.2. W ..........................1 ARNINGS 1.3. A ........................... 1 TTENTIONS 2. OVERVIEW ............................1 2.1. F ..........................2 EATURES 2.2. A ....................
1. Safety 1.1 Instructions for safe operations ◇ Check the main unit and all accessories periodically to make sure that there is no visible damage that may affect patient’s safety and monitoring performance. It is recommended that the device should be inspected once a week at least. Please stop using the monitor when there is obvious damage.
measuring the SpO more easily and accurately, our company developed the Fingertip Pulse Oximeter. The device can measure the pulse rate simultaneously. The Fingertip Pulse Oximeter is tiny, and with low power consumption, convenient to use and carry. You just need to put the fingertip into the sensor of the device, the SpO value will appear on the screen immediately.
transmit some narrow spectrum light bands through blood samples, and to measure attenuation of spectrum with different wavelengths according to the characteristic that RHb, O Hb, Met Hb and COHb absorb the light of different wavelength, thereby determining O Hb saturation of different fractions.
3.3. Clinical restrictions A. As the measure is taken on the basis of arteriole pulse, substantial pulsating blood flow of the testee is required. For a testee with weak pulse due to shock, low ambient/body temperature, major bleeding, or use of vascular contracting drug, the SpO waveform (PLETH) will decrease.
light and that of darkroom is less than ±1%. H. Automatic shutdown function: The device will power off automatically within 5 seconds when there is no finger in the device. I. The view direction (four direction) can be changed by pressing the power button when it works. J.
Step2. Put another end of the cord through the first one and then tighten it. 7. Operating Guide 7.1 Let the patient’s finger put into the rubber cushions of the clip (make sure the finger is in the right position), and then clip the finger. a) Press the Power Button on the front panel.
Page 10
Figure 4 Menu Interface 1) Backlight Adjust Press the button shortly, move to item below, select the “Brightness”, press the button long to adjust the backlight lightness (This product has four mode to adjust the ligntness) 2) Alarm and adjust the up-low limit Press the button shortly, move to the item “Alarm”, press the button long to enter the alarm setting interface as Finger 5: Figure 5 Alarm setting interface...
Page 11
To low the Spo2 and pulse rate limit, put the Dir to Down, press the button shortly to choose the parameter to be adjusted, then press the button long time to low the value. B. Adjust the alarm state Press the button shortly to enter the Alarm item, then long press the button to choose alarm on or off, press “...
Figure 6 Time Setting Menu Adjust the data upload Open the software in the computer, press the button shortly and move to the “ Upload”. Long press the button to choose whether upload the store data in the instrument to the display, if yes, choose “on”.
9. Repairing and Maintenance 1. Please recharge the battery when the LCD screen shows 2. Please clean the surface of the device before using. Wipe the device with alcohol first, and then let it dry in air or clean it by dry clean fabric. 3.
1. The device will power off automatically if no signal is detected 1. Normal. The display is off within 5 seconds. suddenly 2. Recharge battery. 2. The batteries are almost drained. 11. Key of Symbols Symbol Description Type BF Warning – See User Manual The pulse oxygen saturation(%) Pulse rate (bpm) Full-voltage...
Page 15
Measuring range 30bpm~240bpm, (the resolution is 1bpm) Accuracy ±2bpm or±2% (select larger) Pulse Intensity Range Continuous bar-graph display, the higher display indicate the stronger pulse. Battery Requirement Voltage 3.5-4.2 rechargeable lithium battery × 1 Dimensions and Weight Dimensions 57(L) × 32(W) × 30(H) mm Weight About 50g (with the battery)
Need help?
Do you have a question about the CMS-50E and is the answer not in the manual?
Questions and answers