Removal

Refer to Figures 6-1 or 6-2 and 6-3 to identify parts with bold number references.
1. Remove the carrying strap (Section 5.4.1).
2. Loosen and remove the four socket head cap screws 6 (Figure 6-1 or 6-2) from the
base of the contents indicator weight scale mechanism. Remove the contents
indicator assembly 16 (Figure 6-3).

Disassembly

1. While holding the lever 5 down, remove the hitch pin 10 from the end of the spring
rod 9 (Figure 5-5).
2. Remove the lever spacer 13, the spring 11, and the spring rod 9.
3. Use a medium-blade screwdriver to remove the four pan head screws 1 from the
indicator 2. Remove the four screws and the four upper spacers 6 from the indicator
(Figure 5-6).
4. Pry the lens cap 3 away from the base (Figure 5-7).
Figure 5-5
Figure 5-6
Figure 5-7
Service and Repair 5-7