Inspection; Reassembly; Installation - CAIRE COMPANION 1000 Technical Manual

Liquid oxygen portables
Table of Contents

Advertisement

5.5.3
5.5.4
5.5.5
5-8 Service and Repair
5. Remove the indicator needle 4 from the base (Figure 5-8).
6. Use a small screwdriver to remove one E-clip 8 from the pivot shaft 7. Remove the
pivot shaft. Remove the lever 5 (Figure 5-9).
Some older models may contain hitch pins in place of E-clips. If the unit has
hitch pins, it will also contain a different pivot shaft.

Inspection

Inspect all components of the contents indicator weight scale mechanism for any
noticeable signs of wear. Thoroughly inspect the spring rod for signs of excessive wear.
Replace components as necessary.

Reassembly

Reassemble the contents indicator weight scale mechanism by reversing the disassembly
procedure (Section 5.5.2). Lubricate the O-ring in the base of the indicator with a small
amount of Krytox grease. This allows the helix to easily rotate in the indicator.

Installation

Install the contents indicator weight scale mechanism by reversing the removal
procedure (Section 5.5.1). Be sure to put the weight scale mechanism in the proper
orientation. The slotted hole in the mounting bracket should be next to the flow control
valve.
Companion 1000/T Liquid Oxygen System Technical Manual
Figure 5-8
Figure 5-9
B-701964-00 Rev. C

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Companion t

Table of Contents