Unpacking, Installation, And Repacking; Table 1-2: Run Time Specifications; Unpacking; Installation - CAIRE COMPANION 1000 Technical Manual

Liquid oxygen portables
Table of Contents

Advertisement

Companion 1000/T Liquid Oxygen System Technical Manual
1.4.2
1.5
UNPACKING, INSTALLATION, AND REPACKAGING
1.5.1
1.5.2
B-701964-00 Rev. C

Table 1-2: Run Time Specifications*:

Flow Setting
(L/min)
.12
.25
.5
.75
1
1.5
2
2.5
* Specifications subject to change without notice. Data is analytical and based on continuous operation.
Times will vary with ambulation.
Perform the following procedures when unpacking, installing or repackaging a
Companion 1000/T unit.

Unpacking

1. Examine the shipping carton for damage. If the carton is damaged, or its contents
are suspected of being damaged, photograph the damaged carton before the
Companion 1000/T is unpacked. Contact the carrier to request a damage
inspection. Contact the shipping point immediately.
2. Place the shipping carton on a flat surface with the shipping arrows pointing
upwards.
3. Carefully lift the Companion 1000/T out of the carton and remove it from the
plastic bag.
4. Compare the packing list attached to the carton's exterior with the shipment
received. If any discrepancies exist, contact
5. Thoroughly inspect the exterior of the Companion 1000/T for damage (cracks,
etc.).
6. Save all packing materials and the shipping carton for reuse.

Installation

Before installing the Companion 1000/T in a patient's home, read and understand
Section 1.6, Controls, Indicators, and Connectors; Section 1.7, Filling Instructions; and
Section 1.8, Operating Procedures. Perform the following steps upon receipt of
shipment:
Run Time
(hours)
57.1
57.1
21.8
16.7
13.5
9.7
7.6
6.3
Flow Setting
Run Time
(L/min)
(hours)
3
3.5
4
5
6
8
10
15
immediately at 1-800-
Introduction 1-7
5.3
4.6
4.1
3.3
2.8
2.1
1.7
1.1