Inspection; Service; Reassembly - CAIRE COMPANION 1000 Technical Manual

Liquid oxygen portables
Table of Contents

Advertisement

Companion 1000/T Liquid Oxygen System Technical Manual
5.10.3
5.10.4
5.10.5
B-701964-00 Rev. C

Inspection

1. Inspect the Kel-F poppet in the cartridge assembly for wear, damage or embedded
contaminants. Replace the cartridge assembly as required.
2. Inspect the lip seal for cracks, wear, or other damage. Replace the lip seal as
required.
3. Inspect the female adapter for cracks or scratches on the tapered seal surface of the
fill tube connector. Replace the female adapter as required.

Service

Service to the fill connector consists of replacing the cartridge assembly 10, lip seal 12,
female adapter 4, and adapter seal 3.

Reassembly

Cartridge assembly installation requires the use of P/N B-775394-00 female fill
connector installation sleeve (marked "F") and P/N B-775392-00 inner plunger
(marked "I").
1. Insert the cartridge assembly 10 into the female fill connector body 11.
2. Place the larger (lip seal) end of the fill connector flat on a table. Place the stepped
end of the installation sleeve (marked "F") on the smaller (face seal) end of the fill
connector.
3. Place the spiral retainer ring 9 into the open end of the installation sleeve (Figure 5-
25).
Figure 5-24
Figure 5-25
Service and Repair 5-17