Fill Termination - CAIRE COMPANION 1000 Technical Manual

Liquid oxygen portables
Table of Contents

Advertisement

Companion 1000/T Liquid Oxygen System Technical Manual
COM ANION T
2.3.2
ompanion T
B-701964-00 Rev. C
Flow Control
alve
COM ANION 1000
acuum
Gas Molecules
Liquid Oxygen
Gaseous Oxygen
Figure 2-10: Filling

Fill Termination

When the liquid oxygen level in the inner container reaches the end of the vent tube in
the manifold (Figure 2-11), liquid oxygen is expelled through the vent circuit to
atmosphere. When this occurs, the filling operation is terminated by closing the vent
valve and disengaging the Portable from the Stationary unit.
Flow Control
alve
ompanion 1000
acuum
Gas Molecules
Liquid Oxygen
Gaseous Oxygen
Figure 2-11: Fill Termination
Manifold
Warming Coils
ent Tube
ressure
14-16 SI
(96-110 k a)
ent
alve
Manifold
Warming
Coils
ent Tube
Pressure
14-16 PSI
(96-110 kPa)
ent
alve
Secondary
Relief alve
30 PSI
(207 kPa)
22 PSI
(152 kPa)
Primary
Relief
alve
Fill Tube
Liquid
Withdrawal
Tube
Female
Quick Connect
Molecular Sieve
Secondary
Relief alve
30 PSI
(207 kPa)
22 PSI
(152 kPa)
Primary
Relief
alve
Fill
Tube
Liquid
Withdrawal
Tube
Female
Quick Connect
Molecular Sieve
Theory of Operation 2-7