Disassembly - CAIRE COMPANION 1000 Technical Manual

Liquid oxygen portables
Table of Contents

Advertisement

5.10.2
5-16 Service and Repair

Disassembly

1. Hold the female fill connector body 11 with a
the machined flats. Use an adjustable wrench to remove the female adaptor 4
(Figure 5-21).
2. Remove the seal 3 from the female adapter (Figure 5-22).
3. Hold the female fill connector body 11 with a
the machined flats. Use an adjustable wrench to remove the lip seal retainer sleeve
13. Pull the lip seal 12 out of the female quick connect body.
The stepped (spring) end of the lip seal fits into the corresponding recess in the
female quick connect body.
4. Use a small screwdriver or awl to remove the spiral retainer ring 9 by first carefully
lifting the beveled edge of the retainer over the lip of the retaining ring groove
(Figure 5-23).
5. Carefully pry the rest of the ring over the lip until the entire ring pops out.
6. Remove the cartridge assembly 10 (Figure 5-24).
Companion 1000/T Liquid Oxygen System Technical Manual
11
/
-in. open-end wrench placed on
16
Figure 5-21
Figure 5-22
11
/
-in. open-end wrench placed on
16
Figure 5-23
B-701964-00 Rev. C