Verifying Operation; Cleaning The Pulse Oximeter - Nonin 8600 Operator's Manual

Nonin pulse oximeter operator's manual
Table of Contents

Advertisement

Verifying Operation

CAUTION!
The 8600 must be able to measure the pulse properly to obtain accurate SpO 2
measurement. Verify that nothing is hindering the pulse measurement before relying
on the SpO 2 measurement.
Verify that the sensor is properly positioned. Ensure the system is sensing adequate perfusion by observing
that the
indicator is blinking green and correlates to the pulse rate for 10 seconds. Should the perfusion
light be red or yellow or flashing erratically, reposition the sensor or try a different sensor.
For the first 2 minutes of operation, the
disabled during power on. During the first two minutes, the audible alarm may be enabled by pressing the
AUDIO
button. If the audible alarm is not permanently disabled, the
the first 2 minutes of operation and the audible alarm will be enabled.
If the alarm is permanently disabled, the
button will not have no effect on the audible alarm.

Cleaning the Pulse Oximeter

CAUTION!
Do not immerse the 8600 in liquid, and do not use caustic or abrasive cleaning
agents.
The 8600 Pulse Oximeter may be cleaned with a mild detergent and a soft cloth or with an isopropyl alcohol
wipe. Allow enough time for the 8600 to dry thoroughly before reusing.
indicator will flash to indicate the audible alarm is temporarily
indicator will illuminate continuously, and pressing the
8
indicator will not be illuminated after
AUDIO