Specifications
1.
Oxygen Saturation Range (SpO
2.
Pulse Rate Range
3.
Displays
Digital Displays
Patient Indicator
Equipment Indicators
4.
Measurement Wavelengths
Red
Infrared
5.
Accuracy
SpO
(±1 Standard Deviation)*
2
Pulse Rate
6.
Alarm Ranges
High SpO
limit
2
Low SpO
limit
2
High pulse rate limit
Low pulse rate limit
7.
Alarm Volume
8.
Temperature
Operating
Non-operating
9.
Humidity
Operating
Non-operating
10. Battery Charger
Model 8604A
Model 8604A-50
Model 7708
11. Batteries
12. Degree of Protection
13. Dimensions
14. Weight
* Standard Deviation is a statistical measure; up to 32% of the readings may fall outside these limits.
0 to 100%
)
2
18 to 300 pulses per minute
3-digit 7-segment LEDs
Perfusion indicator
Sensor, Self Test, Battery Low, Battery Charging, and Audible Alarm Disabled
indicators
660 nanometers
910 nanometers
70 - 95% ±3 digits for neonates using infant or neonatal sensors
70 - 100% ±2 digits for adults using the Finger Clip Sensor
70 - 100% ±3 digits for adults using Flex or Reflectance Sensors
70 - 100% ±4 digits using Ear Clip Sensor
50 – 70% ±3 digits for adults using Flex or Finger Clip Sensors
Below 50% is not specified for all sensors
±3% ±1 digit
85%, 90%, 92%, 94%, 95%, 96%, 97%, 98%, 99%, OFF
55%, 60%, 65%, 70%, 75%, 80%, 85%, 86%, 87%, 88%,
89%, 90%, 91%, 92%, 95%, OFF
75BPM, 100BPM, 125BPM, 150BPM, 175BPM, 200BPM,
225BPM, 250BPM, 275BPM, OFF
30BPM, 40BPM, 50BPM, 60BPM, 70BPM, 80BPM, 90BPM,
100BPM, 110BPM, OFF
50 to 80 dbA at 1 ft. (30 cm)
0°C to +45°C (+32°F to +113°F)
-20°C to +45°C (-4°F to +113°F)
10 to 90% non-condensing
10 to 95% non-condensing
120 VAC / 60 Hz
220 VAC / 50 Hz
110 VAC / 50 - 400 Hz (meets U.S.A.F. aeromedical requirements)
5-cell rechargeable Ni-Cad battery pack
30 hours minimum operation
15 hours to fully recharge
Type BF Applied Part
5.8" wide x 2.7" high x 7.5" deep
14.7 cm x 6.9 cm x 19 cm
2 lbs (0.75 kg)
21