Activation Instructions - Nonin 8600 Operator's Manual

Nonin pulse oximeter operator's manual
Table of Contents

Advertisement

Features
Controls
All functions of the 8600 are controlled by switches found on the front and rear of the unit. Refer to Figures I
and II for illustrations of these switches.
Power Switch
Pressing the power switch ON (" I ") causes power to be applied to all internal circuitry. Pressing the power
switch to STBY (" O ") causes power to be removed from the displays and puts the pulse oximetry circuitry
into a low power standby mode.
AUDIO Button
AUDIO
The
button has 4 different functions, which are described below.
Function
Audible Alarm Enable/Disable
Calibrate Recorders
Memory Playback (8600M only)
Real-time Setup Mode
Audible Alarm Enable/Disable
Momentarily pressing the
will override the 2 minute period, re-enabling the audible alarm.
NOTE:
Pressing this button has no effect if the audible alarm has been permanently disabled.
Calibrate Recorders
When using a strip-chart recorder for outputting data, calibration between the recorder and the 8600 may be
necessary. To do this during normal operation, hold the
recording output calibrate sequence. You can select the calibration sequence with Option Switch 3.
NOTE:
Calibration signals are sent via the SpO2 and pulse rate lines; not the serial output line.
Standard Calibration:
The value sent for the SpO 2 will alternate from 0% (for 5 seconds) to 100% (for 6 seconds) to 50% (for 5
seconds) and then repeat. The value sent out for the pulse rate will alternate from 0 BPM to 125 BPM to
250 BPM and then repeat. To end calibration sequence, release the
to normal operation.
Alternate Calibration:
The value sent for the SpO 2 will step through at 5% intervals when actuated by the user by pressing the
AUDIO
button beginning at 0% through 100% and then repeat. The value sent out for the pulse rate will
remain at 60 BPM. To end calibration sequence, hold the
return to normal operation.

Activation Instructions

Momentary press during normal operation
Hold for more than 5 seconds during normal operation
Hold for less than 2 seconds during power up
Hold for more than 2 seconds during power up
AUDIO
button disables the audible alarm for 2 minutes. Pressing this button again
AUDIO
button for more than 5 seconds to activate the
AUDIO
button, and the 8600 will return
AUDIO
button for 5 seconds, and the 8600 will
9