Verifying Operation; Startup Sequence And Self-Test - Nonin WristOx2 3150 Operator's Manual

Wrist-worn pulse oximeter with bluetooth wireless technology
Hide thumbs Also See for WristOx2 3150:
Table of Contents

Advertisement

Using the WristOx
Model 3150
2,

Verifying Operation

When the WristOx
Model 3150, first turns on, it performs a startup sequence and self-test. It
2,
occurs:
• When a sensor is applied to a patient (Spot Check mode).
• When a sensor is attached to the device (Sensor Activation mode).
• At a programmed start time when a sensor is attached to the device (Programmed mode).
• After the activation switch is pressed while the device is in Standby mode.
• After the device disconnects from nVISION (Bluetooth connection only).
Verify all indicators display during the startup sequence. Indicators appear in the following order
for 1 second each.

Startup Sequence and Self-Test

1. r and the software revision level:
2. All display icons:
3. Date/time using 24-hour clock format (MM:DD:YY:HH:MM)
(example shows 23 April 2010 at 5:57 p.m.):
Month and Day
(MM:DD)
If the time is not set, the device displays 01:01:10:00:00.
If any indicator does not display, do not use the device. Contact Nonin Technical Service for
assistance.
Year
Hour and Minutes
(YY)
(HH:MM)