Programming Special Functions; Setting The Dosage Concentration And Venting The Dispensing System - Miele G 7883 Operating Instructions Manual

Washer-disinfector
Table of Contents

Advertisement

For future servicing work, please
document any alterations to the stan-
dard factory settings.
Enter any alterations in the field
Value set:_________.
General notes:
– Switch positions that have not been
allocated a function are indicated in
the middle part of the display by a
bar ( - ).
– The programming level is called up
via the STOP position f .
– A number of global parameters and
programme specific parameters
can be altered.
Setting and altering global
parameters (concentration levels:
DOS 1 and DOS 2) applies to all
programmes, in which this
parameter is found.
Setting and altering programme
specific parameters (Wash/rinse
temperatures and temperature
holding times: Main wash 1 and 2;
Final rinse 1 and 2) only applies to
the programme, in which this
parameter is found.

Programming special functions

1. Setting the dosage
concentration and venting the
dispensing system
^ Turn the programme selector to the
STOP position f.
^ Switch off the machine.
^ Press and hold ß and 6 at the
same time, and whilst doing so,
switch the machine on with the I-0
button.
The current programme status "P..."
will appear in the display and the
p / 6 indicator light will come on.
^ Press 0 twice, "E02" (programming
level 2) will appear in the display.