Loading the machine
Note
^ Load items so that water will come
into contact with all surfaces. This
ensures that they will be properly
cleaned.
^ Do not place items to be cleaned
inside other pieces where they may
be concealed.
^ Hollow vessels should be inverted
and placed in the correct inserts. The
water must be able to access them
unhindered.
A cover net can be used to reduce
the risk of movement during the wash
process and prevent glass
breakages.
^ Petri dishes and similar should be
placed in the correct insert with the
soiled side facing the centre.
^ Pipettes should be placed with the
mouthpiece facing upwards.
^ Deep based items should be placed
at an angle to ensure good coverage
of water.
^ Tall, narrow pieces should be placed
in the centre of the baskets This
ensures good coverage of water.
^ Mobile units with adapters must
engage correctly (see next page).
^ The spray arms must not be blocked
by items which are too tall or which
hang down in their path. If necessary,
manually rotate the arms to test.
26
Important:
Make sure that the water connection
spring adapter engages correctly when
a basket or injector unit is inserted in
the machine. It must be 4-5 mm higher
than the water connection inlet in the
machine. If it is not:
^ Loosen the lock ring.
^ Push up the adapter (4-5 mm higher
than the water connection inlet in the
machine) and tighten up the lock
ring.
Select baskets and inserts which are
appropriate for the application.
See examples on the following pages