Programming Special Functions - Miele G 7882 Operating Instructions Manual

Washer disinfector
Table of Contents

Advertisement

For future servicing work, please
document any alterations to the stan-
dard factory settings.
Enter any alterations in the field
"Value set:_________".
General notes:
– Switch positions that have not been
allocated a function are indicated in
the middle part of the display by a
bar ( - ).
– The programming level is called up
via the STOP position f.
– If any changes have been made to
programmes or dispensing systems
these must be documented in a log
book kept with the machine in accor-
dance with the Medical Device Di-
rective (MPG directive 93/42/EWG,
Class IIa) and Health and Safety
Regulations.
1. Setting the dosage concen-
tration and venting the dis-
pensing system
^ Turn the programme selector to the
STOP position f.
^ Switch off the machine.
^ Press and hold ß and 6 at the
same time, and whilst doing so,
switch the machine on with the I-0
button.
The current programme status "P..."
will appear in the display and the p /
6 indicator light will come on.
^ Press 0 twice, "E02" (programming
level 2) will appear in the display.

Programming special functions

1.1 DOS 1 (liquid cleaning agent) via
the optional DOS-Module G 60 or
C 60.
1.1.1 Setting the dosage concentra-
tion:
^ Set the dosage selector on the
DOS-Module to position 10 (applies
to DOS Module C 60 only).
Set the dosage concentration accord-
ing to the cleaning agent manufac-
turer's recommendations.
Example: If the cleaning agent manu-
facturer recommends a concentration
level of 0.5% set the value to 0.50.
^ Turn the programme selector to B.
The display will show either the dos-
age time in seconds for cleaning
agent dispensed via the compart-
ment for powder detergent in the
door, or the dosage concentration for
liquid cleaning agent as a %:
10 = Dosage via the door compart-
ment (10 sec) or
0.10 - 1.0 = 0.1 % -1 % dosage con-
centration.
^ Press ß as many times as neces-
sary or hold it pressed in until the re-
quired value appears.
The dosage time / dosage concentra-
tion is now set.
Value set:
39