Maintenance; Checking Of Measurement Function; Error Displays - Hartmann Tensoval comfort Instructions For Use Manual

Hide thumbs Also See for Tensoval comfort:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 5
any time by pressing the START/STOP
button. Otherwise the device will switch
itself off after a few seconds.
Should the power supply fail, e.g. when
changing the batteries, the stored
values will still be available.
Deleting stored values
You can delete all stored data in M1 or in
M2 for the respective person. To do this,
press the corresponding memory button.
The average value will then appear on the
display. Press down on the memory button
and after four seconds the display will
flash. After eight seconds all data for the
respective person will be deleted and only
M1 or M2 will be displayed. If you release
the button before the end of the eight-
second time limit, no data will be deleted.

9. Maintenance

Do not expose the device to extreme
temperatures, humidity, dust or direct
sunlight.
This device consists of high quality pre-
cision components. Do not drop the
device and protect it against hard
knocks.
The moulded cuff should not be folded
or overstretched.
Never open the device. Repairs may
only be carried out by qualified individu-
als.
Only use a soft, moistened cloth to
clean the device. Do not use detergents
or solvents.
The cuff can be cleaned carefully with
a moist cloth and mild, pH-neutral
soap. Do not completely immerse the
cuff in water.

10. Checking of measurement function

We recommend that measurement func-
tion be checked at intervals of two years
in the case of professional users, such as
pharmacies, medical practices and clinics.
You should also observe the national regu-
lations, such as, in Germany, the Decree
on Sales of Medical Devices. Measurement
function checks can be carried out either
by PAUL HARTMANN AG (see Service
address), by competent authorities, or au-
thorised maintenance providers against
compensation.
Instructions for Calibration Mode:
Remove the batteries in order to switch to
calibration mode. hold down on the
START/STOP button and then insert the
batteries. Next, release the button and,
after a few moments, two zeros will ap-
pear one above each other.
Instructions on checking measurement
function will be supplied on request to
competent authorities, or authorised
maintenance providers.
12
11. Warranty
This product comes with a 3-year war-
ranty. Details of warranty conditions and
contact addresses can be found in the
separate warranty document.

12. Error displays

Error which has occurred
Possible causes
Device will not switch on
No batteries, they have been inserted incorrectly
or are dying. If necessary, check the mains adapter is
connected correctly.
Moving or speaking during the measurement.
Cuff was incorrectly or loosely applied.
Please repeat reading.
Pressure in cuff exceeds 330 mmHg. Therefore, the
pressure will be automatically released.
Batteries are almost dead. Only a few more measure-
ments are possible.
Batteries are dead. Insert new batteries.
English
13

Advertisement

Table of Contents
loading

Table of Contents