i-SENS CareSens N Voice User Manual page 12

Hide thumbs Also See for CareSens N Voice:
Table of Contents

Advertisement

Step 3
Shake the CareSens Control
Solution bottle well before each
test. Remove the cap and squeeze
the bottle to discard the first 1 or 2
drops.
Then wipe the tip with a clean tissue
or cloth. While the
symbol blinks,
apply the control solution to the
narrow edge of the test strip until the meter beeps. Make sure
the confirmation window of test strip fills completely.
Note:
The meter will switch off if the control solution sample
is not applied within 2 minutes of the
the screen. If the meter turns off, press S button or ▲ button
and start from step 2.
Step 4
A test result will appear after the meter counts
down from 5 to 1. If the talking function is on,
the test result will be announced. If you missed
step 2, you can mark the test result as a control
solution test result by pressing and holding the
▲ button for 3 seconds until the ctl appears
on the display. When the ctl is displayed, the
result is stored in the meter's memory but is not
included in the averages.
20 www.i-sens.com
symbol appearing on
Step 5
Compare the result displayed on the meter
to the range printed on the test strip vial.
The result should fall within that range.
Step 6
Slide the test strip ejector button to
discard the used test strip safely in
disposable container.
Caution:
The ranges printed on the test strip vial are for
CareSens Control Solutions only. They do not have any
connection to your blood glucose level.
www.i-sens.com 21