Pre-Cleaning And Disinfection Procedures - i-SENS CareSens N Voice User Manual

Hide thumbs Also See for CareSens N Voice:
Table of Contents

Advertisement

• We have validated Clorox Germicidal Wipes with 0.55%
sodium hypochlorite as the active ingredient for disinfecting
the CareSens N Voice Meter and lancing device. It has been
shown to be safe for use with the meter and lancing device.
This disinfectant is available commercially in towelette form.
In addition to CareSens N Voice BGMS instruction, please read
the instructions provided by the manufacturer of Clorox
Germicidal Wipes before using it.
Name
Clorox® Germicidal Wipes
Manufacturer
Clorox Professional Products Company
[Phone] 1 800 537 1415
[Website] www.cloroxprofessional.com
EPA registration
67619-12
number
Active
Sodium Hypochlorite: 0.55%
ingredients
Note:
The disinfectant products can be purchased online
retailers (e.g. Amazon) or by telephone at the manufacturer.
To find out where to purchase the disinfectant product,
please contact the manufacturer or visit their website as listed
above.
42 www.i-sens.com

Pre-cleaning and Disinfection Procedures:

Open the cap of the
Clorox Germicidal
Wipes container and
pull out 1 towelette
and close the cap.
Wipe the entire surface of the meter 3 times horizontally
and 3 times vertically using one towelette to pre-clean blood
and other body fluids.
Front
Dispose of the used towelette in a trash
bin.
Pull out 1 new towelette
and wipe the entire surface
of the meter 3 times
horizontally and 3 times
vertically using the
new towelette to remove
blood-borne pathogens.
Back
Both sides