i-SENS CareSens N Voice User Manual page 26

Hide thumbs Also See for CareSens N Voice:
Table of Contents

Advertisement

Message
What is said
"Error 5. Strip
This error
insertion
message may
error. Please
appear when the
insert a
wrong blood
CareSens N
glucose test strip
strip properly."
is used instead
of the CareSens
N Single test
strip.
Message
What is said
What it means
"Error 6 / Error 7
There is a
/ Error 8 / Error 9.
problem with the
Meter failure."
meter.
Note:
If the error messages persist, contact Customer Service:
1-800-429-5001 (9 am - 9 pm EST, Monday - Saturday).
48 www.i-sens.com
What it means
What to do
Repeat the test
with a CareSens N
Single test strip.
Insert the strip
with the contact
bars facing up,
and push in
gently until the
meter turns on.
What to do
Do not use the
meter. Contact
Customer Service:
1-800-429-5001.
Message
What is said
What it means
"The room
The temperature
temperature is
during the test is
too low."
below the
operating range.
"The room
The temperature
temperature is
during the test is
too high."
above the
operating range.
What to do
Move to an area
where the
temperature is
within the
operating range
(50-104˚F) and
repeat the test
after the meter
and test strips
have reached a
temperature
within the
operating range.
www.i-sens.com 49