Query for Analog Current Input Level
To query for analog current input level from non-isolated input:
1. Turn the mode control knob to the CAP position or press the Rotary
knob/Enter button if the control knob is already at the CAP position.
CAPr is displayed on the output voltage display.
2. If the mode is currently nonE then select any other APG mode. If this
step is done, be sure to restore the APG setting to nonE when the
operation is complete. Otherwise, do not rotate the Rotary knob/Enter
button to keep the current APG mode.
3. Press the rotary knob/Enter button to select the APG mode.
The output voltage display should read LE C and the output current
display will display APG level.
4. Press the rotary knob/Enter button to return to normal mode.
5. If the APG mode was modified in step 2, execute the procedure to
deactivate the APG voltage mode.
The SCPI command (s) for these instructions are:
Quick Tip
Remote operation
[:]SYSTem[<channel>]:REMote:SOURce:CURRent <?|LOCal
uses these SCPI
|AVOLtage|IAVoltage|ARESistive|IAResistive>
commands. See
page 5–1.
[:]SYSTem[<channel>]:COMMunicate:APRogram:LEVel:CURRent
[:ISOlated]
[:]MEASure[<channel>][:SCALar]:APRogram:CURRent[:DC]?
M370046-01