Analog Programming (APG) and Isolated Analog Programming (ISOL)
Query for ISOL Current Input Level
To query for ISOL current input level from non-isolated input:
1. Turn the mode control knob to the CAP position or press the Rotary
2. If the mode is currently nonE then select any other ISOL mode. If
3. Press the rotary knob/Enter button to select the ISOL mode.
4. Press the rotary knob/Enter button to return to normal mode.
5. If the ISOL mode was modified in step 2, execute the procedure to
The SCPI command for these instructions are:
Quick Tip
Remote operation
[:]SYSTem[<channel>]:REMote:SOURce:CURRent <?|LOCal
uses these SCPI
|AVOLtage|IAVoltage|ARESistive|IAResistive>
commands. See
page 5–1.
[:]SYSTem[<channel>]:COMMunicate:APRogram:LEVel:CURRent
[:ISOlated]
[:]MEASure[<channel>][:SCALar]:APRogram:CURRent
:ISOlated[:DC]?
4-32
knob/Enter button if the control knob is already at the CAP position.
CAPr is displayed on the output voltage display.
this step is done, be sure to restore the ISOL setting to nonE when
the operation is complete. Otherwise, do not rotate the Rotary knob/
Enter button to keep the current ISOL mode.
The output voltage display should read LE C and the output current
display will display ISOL level.
deactivate the ISOL voltage mode.