Disassembly Procedures (A3)

Closed Case Disassembly Procedures

Chapter 6 - Disassembly Guide
This section describes the items that can be removed without disassembling
the main case of the A3 monitor.
Procedure to remove front panel wheel:
The wheel is friction-fit on the encoder shaft.
1. Grasp the sides of wheel firmly and pull straight back from the monitor.
The wheel should slip off the encoder shaft.
2. If wheel does not move, separate front and rear cases as described in
"Separation of Front and Rear Case Assemblies" on page 78. There is a
hole in the front cover behind the wheel. Insert a small screwdriver blade
in the hole and push firmly against the back of the wheel.
Procedure to remove carrying handle:
1. Use screwdriver to remove the two fastening screws and washers. Retain
them for reassembly.
2. Remove the handle by sliding it straight back towards the rear of the
monitor.
Disassembly Procedures (A3)
77