Other Error Codes (All Monitors)

Chapter 5 - Troubleshooting
Hex Code
65, 66
MP-205 SpO
detected an error during
initialization.
6E-71
MP-205 SpO
detected an error on its serial port.
If an error code occurs that is not listed in "Serviceable Hardware Error
Codes", take the following actions:
1. Turn the monitor off, then on again.
2. If the error code still appears, take the monitor out of service and contact
Philips' Response Center or your local Philips representative for advice
on remedial action.
3. If the monitor powers up and the error code does not recur, enter the
Diagnostic Mode and invoke the Error Code screen. Examine the record
of the last 10 error codes and determine if the same error code occurred
previously.
4. If the Error Code screen indicates that the same error has occurred
previously, take the monitor out of service and contact Philips' Response
Center or your local Philips representative for advice on remedial action.
5. If the Error Code screen indicates no previous occurrences of this error,
the monitor can be returned to service.
As a reference, following lists the general categories for other error codes.
The error code categories are shown only in hexadecimal format.
Code (hex)
500xxxx
internal user interface error
501xxxx
remote serial port error
502xxxx
date and time error
503xxxx
NiBP error
504xxxx
front end error
505xxxx
alarm error
506xxxx
audio error
507xxxx
recorder error
Explanation
module has
2
module has
2
Explanation
Troubleshooting Guide
Recommended Action
1. Replace MP-205 module.
2. If problem persists, replace
Main PCB.
1. Replace MP-205 module.
2. If problem persists, replace
Main PCB.