Philips M3921A Service Manual page 21

Patient monitors
Table of Contents

Advertisement

Performance Verification
Table 2
Test and Inspection Reporting
1
Test
Visual: Inspect exterior
of monitor for damage
Power-On Self-Test
Basic Pneumatic
Leakage Test (NBP
test)
NiBP Accuracy Test
NiBP Leakage Test
NiBP Inflation Rate
Test
NiBP Overpressure
Test
NiBP Deflation Rate
Test (A3 only)
Safety Performance
Tests
1. Details of the tests are included later in this chapter.
2. When authorized Philips Medical Systems personnel service the instru-
ment. The results are reported back to Philips. The collected data form a
database to be used in product development. It is not necessary for hospi-
tal personnel to report these results.
3. In the case of the NiBP Deflation Rate Test, applicable to the A3 only.
Expected Results
No Visual Damage
Displays Normal Monitoring
Screen Configuration and emits
tones
After 1 minute at approximately
250 mmHg, pressure drops no
more than 6 mmHg.
250 mmHg; ±5 mmHg
After 1 Minute at 250 mmHg,
pressure drops no more than
6 mmHg
Monitor Reports NiBP = 280
mmHg in <6 seconds
Overpressure Deflation is
activated at a value between
280 mmHg to 330 mmHg
Monitor reports NiBP >10
mmHg and <190 mmHg
All Safety Performance Tests
are in range of Table 5 to
Table 12
Test And Inspection Matrix
2
What to Record
V:P or V:F
PO:P or PO:F
BPL: Px6
Where x6 =Pressure
Drop in mmHg
3
PN:P/X1...X4 (or X5
3
PN:F/X1...X4 (or X5
Record Pressure in
mmHg = X1
Record Pressure
Drop in mmHg = X2
Record Time to
280 mmHg in
Seconds = X3
Record value in
mmHg at which
Overpressure
Deflation occurs =X4
Record value in
mmHg on monitor at
1 min = X5
S:P or S:F
)
)
9