Proper Use
Caution
Note
Chapter 9 - RS-232 Interface
Connection to a battery operated PC should only be
allowed by the end user.
Connection to a line powered device with an ITE Class II
power supply is acceptable provided it passes all 60601-1-1
tests without a problem.
If there is a line powered device with a Class I power
supply, such as a line powered PC or a laser printer in the
system, there must be either optical isolation of the data
cable or the Class I devices must have isolation
transformers.
If you connect the monitor to any instrument, verify proper operation before
clinical use. Refer to the other device's manual for full instructions.
Accessory equipment connected to the monitor's data interface must be
certified according to IEC Standard 950 for data-processing equipment or IEC
Standard 601-1 for electromedical equipment. All combinations of equipment
must be in compliance with IEC Standard 601-1-1 systems requirements.
Anyone who connects additional equipment to the signal inport port or signal
output port configures a medical system and is therefore responsible to ensure
that the system complies with the requirements of system standard IEC
Standard 601-1-1. If in doubt, consult Philips' Response Center or you local
Philips representative.
This ISM device complies with Canadian ICES-001.
Cet appareil ISM est conforme a la norme NMB du
Canada.
The monitor and its accessories must be tested by qualified
service personnel at regular intervals to verify proper operation,
according to the procedures of the user's institution.
Additional important safety information is located in the User's
Guide.