Hardware and Software Tests

Hardware and Software Tests
SpO
Testing
2
Alarms and Alarm
Silence
14
Hardware and software testing includes the following tests applicable to the
indicated models in the series:
Test
SpO
Testing
2
Operation with an ECG Simulator
Operation with a Respiration Simulator
Verification of Pneumatic System
Operation with a Temperature Simulator
General Operation
SpO
testing includes the following tests:
2
Alarms and Alarm Silence
n
Heart Rate Tone Volume Control
n
Dynamic Operating Range
n
LED Excitation Test
n
1.
Connect the SRC-2 pulse oximeter tester to the M4787A sensor cable and
connect the cable to the monitor.
2.
Set SRC-2 as follows:
SWITCH
RATE
LIGHT
MODULATION
RCAL/MODE
3.
Press the ON/STANDBY
4.
After the normal power-up sequence, verify that the SpO
Applies to Model(s)
A1, A3 - all models
A1 - M3923A, M3924A
A3 - all models
A3 - all models
A1, A3 - all models
A1 - M3922A, M3924A
A3 - all models
A1, A3 - all models
POSITION
38
LOW
OFF
RCAL 63/LOCAL
button to turn the monitor on.
2
Performance Verification
% display initially