Analog Output Signal Connections
© National Instruments Corporation
The analog output signals are DAC0OUT, DAC1OUT, EXTREF, and
AOGND. DAC0OUT and DAC1OUT are not available on the
PCI-6032E and PCI-6033E. EXTREF is not available on the
PCI-MIO-16XE-10, PCI-MIO-16XE-50, PCI-6031E, PCI-6032E, or
the PCI-6033E.
DAC0OUT is the voltage output signal for analog output channel 0.
DAC1OUT is the voltage output signal for analog output channel 1.
EXTREF is the external reference input for both analog output
channels. You must configure each analog output channel individually
for external reference selection in order for the signal applied at the
external reference input to be used by that channel. If you do not specify
an external reference, the channel will use the internal reference.
Analog output configuration options are explained in the Analog Output
section in Chapter 3, Hardware Overview. The following ranges and
ratings apply to the EXTREF input:
•
Usable input voltage range: ±11 V peak with respect to AOGND
•
Absolute maximum ratings: ±15 V peak with respect to AOGND
AOGND is the ground reference signal for both analog output channels
and the external reference signal.
4-23
Chapter 4
Signal Connections
PCI E Series User Manual